Reports until 15:06, Friday 13 May 2016
H1 ISC
nutsinee.kijbunchoo@LIGO.ORG - posted 15:06, Friday 13 May 2016 - last comment - 16:03, Monday 16 May 2016(27183)
PSL power precisions measurement

Quick conclusion:

Bad news: The further away the rotation stage has to travel, the less accurate the measured requested power becomes.

Good news: The uncertainty is within 1-3%.

 

Slightly more details:

Images attached to this report
Comments related to this report
nutsinee.kijbunchoo@LIGO.ORG - 16:03, Monday 16 May 2016 (27229)

A little further investigation shows that the power deviation during the second set of measurements started at 58th measurement where 19 W jumped to nearly 20 W and 54 W jumped to 56 W (first attachment). This jump did not correspond to the measured angles which stayed consistant throughout the measurement (second attachment). Third attachment shows the histogram of power of the first set of measurement (going back and forth between 0W and 3W) that didn't appear to have any power jump.

Images attached to this comment