Displaying report 1-1 of 1.
Reports until 16:43, Thursday 19 May 2016
H1 TCS
nutsinee.kijbunchoo@LIGO.ORG - posted 16:43, Thursday 19 May 2016 (27308)
Quick visual inspection of TCS AOMs

Aidan+Alastair (on phone), Bryn, Nutsinee

WP#5881

 

Quick conclusion:

We are starting to have the same nasty particulates problem that LLO was having up until late April (I assume the problem was somewhat solved there. See LLOalog25658  and many more alogs after this one by Alastair). CO2X is in a better shape (less particulate clumps) compared to CO2Y.

 

Details:

 

We're going to need a proper procedure on how to resolve this issue. This cannot, and will not, go on until O2.

Both CO2X and Y were restored after the inspection but I don't really like the idea of leaving chillers running after finding these particulates.

Also we don't use DI water for the chillers. We use the same water our PSL is using.

Images attached to this report
Displaying report 1-1 of 1.