GregG & HughR Greg went in chamber and added the screws to the ISC Cables that had none. These were reconnected and are now secure to the FeedThru. These were the ISC Transmon 532nm & Beam Diverter cables. Also, Greg connected the one still dangling cable to F3-3C2 (D1003085), presumably this is the TCS-Ring Heater cable. Next we took some level shots on the Optical Table and saw +-0.2mm level. We made some slight HEPI tweaks and using the dial indicators as a guide to what we did, we now have the level to +-0.1mm. So this should be good. We'll make the Actuator Connections next.
Note my earlier alog entry.
https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=2586