Reports until 14:24, Friday 03 June 2016
H1 INJ (INJ)
keith.riles@LIGO.ORG - posted 14:24, Friday 03 June 2016 - last comment - 20:04, Tuesday 07 June 2016(27548)
Implemented new inverse actuation for CW injections with time delay correction
I have copied Evan's new actuation function to the h1hwinj1 directory currently used
for CW injections: ~hinj/Details/pulsar/O2test/. I used the one that corrects for the
actuation delay:  H1PCALXactuationfunction_withDelay.txt.

For reference, the uncorrected version (no "_withDelay") sits in the same directory,
along with the one we first tried last week: H1PCALXactuationfunction.txt.25may2016.
The perl script that generates the command files in.N (N=0-14) has been updated to
use "_withDelay" and the command files regenerated.

The CW injections have been killed and automatically restarted by monit. Attached 
are second trends before and after the gap showing that things look about the same,
as expected, but there is a small increase in injection amplitude (~5%).
Images attached to this report
Comments related to this report
keith.riles@LIGO.ORG - 20:04, Tuesday 07 June 2016 (27642)INJ
Evan wondered if the ~5% increase in total injection amplitude was dominated
by the highest frequency injection or one at lower frequencies. I took a look 
for this time interval and found that the total amplitude is dominated by
the injection at ~1220.5 Hz. Simply comparing spectral line strengths 
before and after the changeover turned out not to be a robust way to
estimate the frequency-dependent ratio of the new to the old inverse actuation 
function, because some pulsar injections (especially the highest frequency one)
are going through rapid antenna pattern modulations during this period.

But comparing the new to the old spectral line strengths at the same
sidereal time several days later (after power outage recovery) gives
robust measures for a sampling of low-, medium- and high-frequency
injections:

Freq (Hz)"Old" amplitude (before switchover)New amplitude (4 sidereal days later)Ratio (new/old)
190.900.322920.322111.00
849.0060.50262.3441.03
1220.5299.37318.701.06
1393.1207.50224.371.08
1991.228.56532.7881.15
These results seem in reasonable agreement with Evan's expectation of a new/old ratio rising with frequency, reaching 15% at 2 kHz. Plots and text files of 4-minute spectral averages are attached for June 3 before the switchover and for June 7 with the newer inverse acuation function.