chandra.romel@LIGO.ORG - posted 14:48, Tuesday 14 June 2016 (27726)
Pirani gauge pot adjustment
Gerardo, Chandra
Adjusted the potentiometers of the following pirani gauges (3-8 turns in CW direction) to calibrate pressure reading to ~5e-4 Torr.
PT-100
PT-110
PT-114
PT-134
PT-210