Reports until 16:15, Tuesday 21 June 2016
H1 PSL
keita.kawabe@LIGO.ORG - posted 16:15, Tuesday 21 June 2016 - last comment - 18:56, Tuesday 21 June 2016(27895)
ISS board transfer function

I injected into the 2nd loop board via PSL-ISS_TR (3rd looop filter) and SR560, and measured the transfer function from the readback of the analog input (H1:PSL-ISS_SECONDLOOP_PD_58_SUM_OUT, which is 32k channel) to the readback of the analog output ( H1:PSL-ISS_SECONDLOOP_SIGNAL_OUTPUT, which I think is in the 1st loop board) when the gain is set to 0, BOOST off, integrator off.

I was confused because of an additional 500Hz pole, but it seems like this was a questionable 500Hz digital pole in H1:PSL-ISS_SECONDLOOP_SIGNAL FM1. Anyway, in the attached, dots are as measured but after counter-compensating for the questionable 500Hz digital pole, and the lines are what is expected from D1300439 and S1400214.

The analog TF of the board itself should be something like zpk([1, 100, 8.46k, 10k], [0.1, 0.5, 500, 33.9k], 350). Yes there's 500Hz pole in analog. This is like one 0.1Hz pole up to 100Hz with some funny additional things (but don't forget SR560 with AC coupling and 10Hz pole, SR560 is used as a poor man's dewhitening to cut the DAC noise).

To make the 3rd loop design without 2nd loop somewhat easier, I made a simple zpk([0.1;0.5], [1;10; 50], 1) in FM9 of H1:PSL-ISS_TR called boardComp to get rid of p=[0.1, 0.5] and z=1. The potential problem of this is that the SR560 might rail. If this happens, we need something better.

Anyway, the starting point is to turn this filter on, then set the gain to whatever it was originally and divide it by 350, and measure the OLTF of the third loop without the second loop while the loop is open.

Images attached to this report
Comments related to this report
keita.kawabe@LIGO.ORG - 18:27, Tuesday 21 June 2016 (27898)

And it worked (Sheila, Kiwamu, Stefan, me).

No 2nd loop, boost off, integrator off, 20dB gain off.

Attached MEDM screen shows the filters used, and the dtt session shows the open loop transfer function when the loop gain was -1. At 2W, we were able to go as high as -3.

However, at 11W, when we turned the loop on with the gain of -3, something started oscillating with about 30 seconds period and we couldn't save the lock.

Images attached to this comment
keita.kawabe@LIGO.ORG - 18:56, Tuesday 21 June 2016 (27900)

Questionable 500Hz pole in H1:PSL-ISS_SECONDLOOP_SIGNAL FM1 was removed.