/data file system on Hartmann Wavefront Sensing computer is full

Nutsinee, Jim, Dave:

The HWS code crashed at 07:50 PDT this morning, Nutsinee tried to restart at 11:46 PDT, but it failed. We found that the 1TB raid is 100% full (has data back to December 2014). We are contacting Aidan to see how to proceed.

 

BTW: the /data file system on h1hwsmsr is NFS mounted at the end stations, so no HWS camera information is being recorded at the moment.

Comments related to this report
david.barker@LIGO.ORG - 13:44, Thursday 23 June 2016 (27938)

We deleted December 2014 ITMX and ITMY data to free up 21GB of disk space on /data. The code now runs.

We need a long term plan on how to keep these data if they need permanent archiving.

kiwamu.izumi@LIGO.ORG - 19:38, Monday 27 June 2016 (27991)

I have restarted both X and Y HWS codes this evening.

nutsinee.kijbunchoo@LIGO.ORG - 20:20, Tuesday 28 June 2016 (28038)