SUMMARY: Had the Control Room to myself in the evening. Unfortunately, H1 was a bit finicky (see Locking & Initial Alignment notes below) & drove me crazy for an hour or two, but eventually it went to 20W. Will probably leave it here for the night. Violin modes (& harmonics) are rung up.
Config Changes noted for EY on the CDS Overview:
ISI: If we want to address this with reboot, take ISI to a non-sensor correction state first. Then load new filters. But can leave until the morning for Jim to take care of this.
Commissioning:
Soft Loop investigations at the beginning of the shift. (Jenne, Sheila)
Locking Notes:
Switched the input matrix values back for the TMS Soft Loops which Sheila alogged about earlier. Tried locking, but DRMI did not look good.
Initial Alignment Notes:
Went through a couple of alignments and both times had issues with INPUT_ALIGN. For the first time, the MC would not hold lock (so I skipped it). For the second time, the Xarm flashes were good, but it would not lock. I doubled the XARM gain, but I couldn't drop it back down without the Xarm breaking lock (aligned while keeping the gain at the doubled value of ~0.15). Then moved on.
After the first Initial Alignment, could not make it past LOCKING_ALS. Eventually went for the 2nd Initial Alignment (& had INPUT_ALIGN issue noted above, but atleast made it through the step). Went for locking and this time had none of the LOCKING_ALS issues!
[Puzzled about what the deal is with INPUT_ALIGN! Atleast in my attempts, it seemed to definitely affect whether I could get past LOCKING_ALS.]
Leaving H1 at the INCREASE_POWER (@20W) Guardian state for the night.