Displaying report 1-1 of 1.
Reports until 16:00, Monday 27 June 2016
H1 SUS (SUS)
haocun.yu@LIGO.ORG - posted 16:00, Monday 27 June 2016 (27985)
Check for switching ESD Drivers to lower range

We plan to change to use the ETMY ESD drivers with lower range and lower noise. To make sure that this will not saturate the drivers, we measured the spectrum of ETMX, and calibrated it with ESD OUTPUT filters, which are different for ETMX and ETMY. This should be fine because the calibrated magnitude is lower than 1000 cnts, which is much smaller than the driver's range (~130000 cnts).

Images attached to this report
Displaying report 1-1 of 1.