Displaying report 1-1 of 1.
Reports until 16:35, Monday 27 June 2016
H1 PSL
patrick.thomas@LIGO.ORG - posted 16:35, Monday 27 June 2016 - last comment - 17:44, Monday 27 June 2016(27988)
Ops PSL Weekly Report
The following was generated from running /opt/rtcds/userapps/release/psl/h1/scripts/PSLweekly.py. It appears the nominal values may need updating? Note that the wiki says to run /opt/rtcds/userapps/release/psl/common/scripts/PSLweekly.py, but that script reports errors. TJ pointed me to this one.

Laser Status:
SysStat is good
Front End Power is 30.38W (should be around 30 W)
Front End Watch is GREEN
HPO Watch is GREEN
PMC:
It has been locked 6.0 days, 5.0 hr 8.0 minutes (should be days/weeks)
Reflected power is 30.07Watts and PowerSum = 147.2Watts.

FSS:
It has been locked for 0.0 days 0.0 hr and 17.0 min (should be days/weeks)
TPD[V] = 3.834V (min 0.9V)

ISS:
The diffracted power is around 15.37% (should be 5-9%)
Last saturation event was 0.0 days 0.0 hours and 17.0 minutes ago (should be days/weeks)

Comments related to this report
thomas.shaffer@LIGO.ORG - 17:44, Monday 27 June 2016 (27989)

There are two copies of this script for some reason, not sure if I just wasn't paying attention or it just got accidentally copied. I made sure that the one in /psl/common/scripts/ is the most up to date and also the one aliased into ops, and then removed the one living in h1 from local and svn.