patrick.thomas@LIGO.ORG - posted 10:21, Thursday 30 June 2016 - last comment - 13:42, Thursday 30 June 2016(28083)
h0vacmx Beckhoff vacuum controls updated
WP 5972

The Beckhoff vacuum controls code on h0vacmx has been updated to use the same PI controller as h0vacey (alog 27886).

I have put both CP5 and CP6 on PID control.

The channels in the DAQ and autoBurt request snapshots still need to be updated.
Comments related to this report
patrick.thomas@LIGO.ORG - 13:42, Thursday 30 June 2016 (28091)
CP5 and CP6 settings and response attached. The initial increase in the CP6 pump level happened because I first had the minimum output set at 50, which was too high. I lowered it to 20 and it recovered.