biggest = prm pitch changing by 2.5urad
close second = im4 pitch changing by 1,15urad
plot and chart of top 16 optics/DOFs attached
plot of optics/DOFs in the chart - meant to attach yesterday. I
used OpLev signals when available and OSEMs when not.
UPDATE to my alog's less than descriptive title: I think the original title suggests these changes were during a 40W lock, but this is not the case.
The optic alignment changes are actually happening over 8 minutes of full lock while the input power increased from 22W, the O1 power level, to 40W.