H1 PSL (PSL)
peter.king@LIGO.ORG - posted 07:09, Tuesday 05 July 2016 - last comment - 09:11, Tuesday 05 July 2016(28142)
Laser down
Came in to find the laser was off due to the power watchdog being tripped.
Both chillers were still running.

I have turned the laser back on but it hasn't returned to where it normally
returns to.  Looking into what happened.
Comments related to this report
peter.king@LIGO.ORG - 07:12, Tuesday 05 July 2016 (28143)
The laser tripped around 4:15 this morning.
peter.king@LIGO.ORG - 09:11, Tuesday 05 July 2016 (28144)
Looks like the NPRO tripped out.  Maybe due to a minor power surge due to the start of the
week?

    Pre-modecleaner had some problems locking irrespective of the reference level.  Found
that a small tweak to the horizontal alignment brought things back right away.

    Note in passing: with no light the output of the locking photodiode was ~ +29 mV.
Unlocked (ie with light incident on the pre-modecleaner) the output was ~ -230 mV.
Which is sort of low, which might be consistent with a small mis-alignment.  The
locking photodiode is a 1 mm InGaAs photodiode after all.