Attached are measurements of the ISS transfer function performed after a
minor alignment adjustment to the ISS AOM. These were taken with the gain
slider at 6 dB and 9 dB. The offset slider was left at 6.35.
With the gain slider at 6dB the UGF is about 54 kHz; 9 dB 70 kHz.
However there is considerable peaking in the noise spectrum. The free-running
noise is also attached.
All measurements were taken with PDA which was the in-the-loop sensor
at the time. The reference signal was at -2.00. At this setting the measured
DC output of PDA was -8.70 V.
Jason/Peter