Displaying report 1-1 of 1.
Reports until 05:35, Thursday 07 July 2016
H1 PSL (PSL)
peter.king@LIGO.ORG - posted 05:35, Thursday 07 July 2016 (28218)
PMC visibility
With the pre-modecleaner unlocked the output of the locking photodiode was between
-1.070 V and -1.080 v.  Locked was between -170 mV and -185 mV.  This gives the
pre-modecleaner visibility to be between 83% and 84%.


Jason/Peter
Displaying report 1-1 of 1.