State of H1: locked at 25W, but no ISS 2nd loop
Activities:
-
Peter King - PSL - FSS tuning
-
Kyle - MX - vacuume work
-
Robert - electronics room
-
JimW - HAM2 blend filter update
Glitches:
-
PSL AC left on - now off
-
Confusion over a guardian state - lock loss
-
ISS 2nd loop is not working - currently being worked on
Special Instructions:
-
SRM human servo - found on ASC medm tab, SRM POP90 - use to keep the signal low and smooth (offsets in guardian were for 40W, but Sheila fixed by taking them out)
After one lock loss FSS started oscillating (PZT signal H1:PSL-FSS_FAST_MON_OUTPUT was going rail to rail).
Bringing down the fast gain by 1dB (from 22 to 21) quenched it, and though it didn't start oscillation when the fast gain was brought back up 1dB, I just leave it at 21.
Update:
FSS oscillated multiple times after the above was written.
The temporary cure seemed to be to lower the FSS common gain and then bring it back up. No need to change FAST gain.