Laser Status:
SysStat is good
Front End power is 34.25W (should be around 30 W)
Frontend Watch is GREEN
HPO Watch is GREEN
PMC:
It has been locked 0.0 days, 1.0 hr 0.0 minutes (should be days/weeks)
Reflected power is 22.93Watts and PowerSum = 152.9Watts.
FSS:
It has been locked for 0.0 days 0.0 h and 27.0 min (should be days/weeks)
TPD[V] = 3.377V (min 0.9V)
ISS:
The diffracted power is around 14.75% (should be 5-9%)
Last saturation event was 0.0 days 0.0 hours and 27.0 minutes ago (should be days/weeks)