Finally, I was able to get the Testo ComSoft5 software downloaded in order to look at the relative humidity logs for the desiccant cabinets in the clean and bake warehouse. Attached are the plots for the SUS DB1 cabinet and the 3IFO DB4 cabinet. The trending software plots back to the last time the data was plotted which in this case was last done in June. I am glad to see that there is a bump in the RH just after the power outage in June since prior months plots are usually ~flatline and suspect. Also for the last 6 months, one of the RH meters has been hanging out near -0.4% RH which we need to get to the bottom of, but at least the meter reads positive ~20-40% RH when I move it outside, and during the last power outage.
Past monthly plots are in alogs 23049 22277 19413 18826 17611
FYI:
The data logger software ComSoft5 is now loaded on the Contamination Control HP Mini laptop which resides in a bin outside of my office.
Data is in the file Documents/Desiccant Data/
In the bin are the cables and RH meter boxes/manuals.
The keys to the des cabinets (which are needed to pull out the little RH meter to plug into the computer for data retrieval) are in a variety of places (a central spot is TBD) - however it turns out that all keys work on all 4 cabinet locks in the C&B warehouse.