Reports until 11:26, Thursday 21 July 2016
H1 PSL (PSL)
peter.king@LIGO.ORG - posted 11:26, Thursday 21 July 2016 (28555)
Laser shutdown
Travis called, said the laser had just tripped out.

    From the status screen it looks like it was a flow rate error in the power meter cooling
circuit - the same suspected reason as last time.  This in turn trips the power watchdog.
Hopefully this is a sensor problem and not one of hardening of the arteries.

    Note the laser shutter(s) remain open.

   The injection relock counter was reset to 0 on both the Beckhoff PC and the MEDM screen.
Images attached to this report