I am working on the Noisebudget Model these days and made a summary of the WFS in use with calibrations:
| WFS |
Serial No. |
RF Trans. [Ω = V/A] |
Responsivity [A/W] |
Demod. Gain [dB] |
Whit. Gain [dB] (40W) |
ADC [cts/V] |
Digital Gain |
Cal [cts/W] |
|||||
|
Freq. |
Q1 |
Q2 |
Q3 |
Q4 |
AVE |
||||||||
|
REFL_A |
S1301242 |
9 MHz |
894 |
894 |
1069 |
974 |
958 |
0.85 |
21 |
21 |
1638 |
1 |
1.7E8 |
|
45 MHz |
642 |
662 |
657 |
634 |
649 |
21 |
1 |
1.14E8 |
|||||
|
REFL_B |
S1301243 |
9 MHz |
822 |
873 |
742 |
861 |
824.5 |
0.85 |
21 |
21 |
1638 |
1 |
1.44E8 |
|
45 MHz |
678 |
644 |
635 |
648 |
651 |
21 |
1 |
1.14E8 |
|||||
|
AS_A |
S1300635 |
36 MHz |
810 |
828 |
872 |
891 |
850 |
0.85 |
21 |
12 |
1638 |
1 |
5.3E7 |
|
45 MHz |
707 |
718 |
732 |
766 |
731 |
3 |
1 |
1.6E7 |
|||||
|
AS_B |
S1300634 |
36 MHz |
846 |
904 |
826 |
832 |
852 |
0.85 |
21 |
15 |
1638 |
1 |
7.5E7 |
|
45 MHz |
644 |
711 |
719 |
691 |
691 |
3 |
1 |
1.5E7 |
|||||
Calibration (from W to cts) = RF Trans. * ADC * Responsivity * Digital Gain * 10^(Demod. Gain + Whit. Gain)/20
Just a note for ASC WFS REFL_A: S1301242 was swapped out in 2022 for S1301248. See alog #63544.