ITM oplevs need more whitening gain, ITMY oplev needs more whitening filters, we could also give ETMX one more whitening filter.

According to T1500556, current settings for test mass oplevs are:

  Whitening gain Number of active whitening filters oplev SUM counts
ETMX 18dB 1 40500
ETMY 0dB 2 28000
ITMX 0dB 2 2400
ITMY 0dB 0 5400

ITMY oplev PIT noise is close to the DAQ noise floor at about 3Hz, and the DAQ noise dominates the RMS down to 0.7Hz or so. ITMX is OK for RMS due to double 1:10 whitening, but not much better at 3Hz. Not good for oplev damping.

Also it seems totally fine to use two stages of whitening filters for ETMX.

We can set things up like this (changes in bold font):

  Whitening gain Number of active whitening filters 4000*10^(24/20)oplev SUM counts
ETMX 18dB 2 40500
ETMY 0dB 2 28000
ITMX 21dB 2 27000
ITMY 15dB 2 30000
Images attached to this report