According to T1500556, current settings for test mass oplevs are:
Whitening gain | Number of active whitening filters | oplev SUM counts | |
ETMX | 18dB | 1 | 40500 |
ETMY | 0dB | 2 | 28000 |
ITMX | 0dB | 2 | 2400 |
ITMY | 0dB | 0 | 5400 |
ITMY oplev PIT noise is close to the DAQ noise floor at about 3Hz, and the DAQ noise dominates the RMS down to 0.7Hz or so. ITMX is OK for RMS due to double 1:10 whitening, but not much better at 3Hz. Not good for oplev damping.
Also it seems totally fine to use two stages of whitening filters for ETMX.
We can set things up like this (changes in bold font):
Whitening gain | Number of active whitening filters | 4000*10^(24/20)oplev SUM counts | |
ETMX | 18dB | 2 | 40500 |
ETMY | 0dB | 2 | 28000 |
ITMX | 21dB | 2 | 27000 |
ITMY | 15dB | 2 | 30000 |