[Betsy, Koji]
We brought the OMC to the bonding lab for inspection. We found some of the EP30-2 joints between the main glass breadboard and the invar mounting brackets showed sign of delamination. Dennis, Calum, and Garilyn have been notified and are woking on the counteraction.
Attachment 1: 40% delamination. This piece is holding one side of a balance mass holding bracket.
Attachment 2: 80% delamination. This piece is holding one side of a balance mass holding bracket.
Attachment 3: 40% delamination. This piece is holding one side of a balance mass holding bracket.
Attachment 4: 80% delamination and this piece is supporting one of the DCPD housing at the bottom side.
Attachment 5: 30% delamination and this piece is supporting the other DCPD housing at the bottom side.