Displaying report 1-1 of 1.
Reports until 12:52, Wednesday 03 August 2016
H1 ISC (ISC)
corey.gray@LIGO.ORG - posted 12:52, Wednesday 03 August 2016 - last comment - 10:08, Tuesday 09 August 2016(28842)
HAM6 Photos From Yesterday (8/2): Damage, Dust, & Action Shots

(Betsy, Corey, Keita, Koji & also some photos by Richard M)

Following up on alogs from Keita & Koji, photos from yesterday's work has been uploaded to resourcespace and can be found here:

https://ligoimages.mit.edu/?c=1696

(There's 91 photos.  I did not upload blurry photos.)

I'm attaching highlight photos to this alog & they are roughly numbered in the order taken:

  1. M14 (was removed to blow particle off)
  2. Fingerprints on NW corner of shroud
  3. V-beam Dump (on/near QPD sled on northern part of table) with major damage on both pieces of glass
  4. Burn hole at entrance aperature on the shroud glass for OMC
  5. Photos of damage on V-Beam Dump after pulled out and on inspection table.
  6. Photos of damage on V-Beam Dump after pulled out and on inspection table.
  7. Photos of damage on V-Beam Dump after pulled out and on inspection table.
  8. Action shots of Betsy/Koji & OMC removal
  9. Action shots of Betsy/Koji & OMC removal
  10. Action shots of Betsy/Koji & OMC removal
  11. OMC PZT mirror with burn mark
  12. OMC PZT mirror with burn mark
  13. Particle contaminants on OMC Breadboard
  14. Delaminated epoxy on one of the "tomb stones"
Images attached to this report
Comments related to this report
koji.arai@LIGO.ORG - 20:24, Wednesday 03 August 2016 (28861)

> 4. Burn hole at entrance aperature on the shroud glass for OMC

Today, we took out the black glass panel with a burn hole from the OMC shroud.

The attached close-up photo shows the burn hole above the entrance aperture to the OMC. Also, you can observe many more craters.

As we are going to install the replacement panel, we wonder what was the cause of these misaligned craters and what was the difference between the small craters and the biggest burn hole.

Images attached to this comment
betsy.weaver@LIGO.ORG - 10:08, Tuesday 09 August 2016 (28952)

~10:00AM WED AUG 4th, 2016 - Chamber entrance Particle counts with hand held CC counter:

Particle size Outside of chamber cover, in CR In chamber above table
0.3um 70 0
0.5um 20 0
1.0um 10 0

 

~1PM WED AUG 4th, 2016 - Chamber entrance Particle counts with hand held CC counter:

Particle size Outside of chamber cover, in CR In chamber above table
0.3um   0
0.5um   0
1.0um   10
Displaying report 1-1 of 1.