Results from testing Y2-8 Ion Pump HV Cable (update found the short)
Gerardo, Marc

We tested the HV Ion Pump cable at the Y end station with a DMM and the Fieldfox this morning.  After pulling the cable out of the conduit we found damage on the cable about 17m from the connector.  The damage is electrical in nature and is about where the Fieldfox predicted, see image. 