H1 PSL
edmond.merilh@LIGO.ORG - posted 13:31, Wednesday 03 August 2016 - last comment - 08:48, Thursday 04 August 2016(28846)
DBB Scans FAMIS# 6584

RPN: 35W- no change

        200W- no change    

FRQ: 35W- no change from 10Hz to 1K. Control signal is better than last week from 1-10Hz. Error signal higher at HF.

        200W- no change

PNT: 35W- no change

         200W- no changes to speak of. High peaks at 100Hz, 1K and HF noise.

MSC: 35W-  no change. TEM10 looks a bit high. HOM power and cts are down marginally from last week

        200W- no significant change. Plot actually looks slightly better than last week despite no adjustments. Errors during measurement can be seen in the .pdf below.

Images attached to this report
Non-image files attached to this report
Comments related to this report
jason.oberling@LIGO.ORG - 08:48, Thursday 04 August 2016 (28866)

Ed's analysis is spot on.  No serious changes from previous scans.