H1 ISC
koji.arai@LIGO.ORG - posted 18:10, Friday 05 August 2016 - last comment - 09:51, Tuesday 09 August 2016(28912)
HAM6 OMC work progress

[Corey, TJ, Betsy, Koji]

RESOLVED ISSUES:
- The issue of the suspension TF was resolved by adjusting one of the OMC EQ stop holders [LHO ALOG 28900]

- The sign-flip issue of the DCPD signals has been explained by the accoustic noise of the OMC [LHO ALOG 28901]

PROGRESS:
- The lock nuts of the upper mass EQ stops were fastened (by fingers)
- The black glass panels of the OMC shroud were restored. We checked the input/reflection apertures to be well aligned against the beams.
- The OMCT steering mirror was restored to the optical mount.

- The alignment of the WFS path was reviewed. It was found that the path is still well aligned. Just the WFS QPDs were manually aligned by the mounts with picomotors.

- We attached the viewport simulator on the flange to check the alignment of the in-air paths. The AS AIR path was very well centered on the viewport without touching. The OMCR path is about 50% towards East (-Y) from the center of the viewport. We could not find the OMCT spot no matter how we flash the OMC. This could be just the beam was too dim, or was blocked by the shroud. We'll investigate this on Monday again.


TO DO:

- Continue to work on the OMCT path alignment. This requires ~10W IMC input. Try to find a dim flashes. Use a CCD camera on a tripod to find the spot around the viewport cover.
- We want to check the calibration between AS_C QPD SUM, the incident power on OM1, and the incident power on the OMC breadboard.
- Other ISC items function check (picos, beam diverters, fast shutter - see Rich's comment below, etc)
- Ground loop check
- Other SUS/SEI exit check

Comments related to this report
jeffrey.kissel@LIGO.ORG - 18:23, Friday 05 August 2016 (28913)DetChar, ISC, SUS
J. Kissel

Now that the re-assembly of the OMC is complete including EQ stop spacing, EQ stop lock-nut securing, and OMC shroud replacement, I've reconfirmed that the dynamics of the OMC SUS are still free-and-clear as they were this afternoon. Nice work Koji and co.! There is no further planned installation work in and around the OMC SUS, so we suspect these will be the last full in-air transfer functions before chamber close-out next week.

Templates live here:
/ligo/svncommon/SusSVN/sus/trunk/OMCS/H1/OMC/SAGM1/Data/
2016-08-06_0007_H1SUSOMC_M1_WhiteNoise_L_0p02to50Hz.xml
2016-08-06_0007_H1SUSOMC_M1_WhiteNoise_P_0p02to50Hz.xml
2016-08-06_0007_H1SUSOMC_M1_WhiteNoise_R_0p02to50Hz.xml
2016-08-06_0007_H1SUSOMC_M1_WhiteNoise_T_0p02to50Hz.xml
2016-08-06_0007_H1SUSOMC_M1_WhiteNoise_V_0p02to50Hz.xml
2016-08-06_0007_H1SUSOMC_M1_WhiteNoise_Y_0p02to50Hz.xml

Note, I've finally listened to my own advise and made sure that these in-air transfer function templates all have the same frequency resolution and excitation amplitude and coloring. 
Images attached to this comment
rich.abbott@LIGO.ORG - 18:38, Friday 05 August 2016 (28914)ISC
While HAM6 is open, it would be good to do a close inspection of the fast shutter.  Specifically the wires attaching the bobbin (toast) to the integral terminal block.  These wires are tiny and are the most likely item to fail.  Also good to check for any delamination of the mirror to the bobbin.
corey.gray@LIGO.ORG - 08:44, Monday 08 August 2016 (28922)

Photos from Friday's work is on Resource Space here:

https://ligoimages.mit.edu/?c=1703

Images attached to this comment
betsy.weaver@LIGO.ORG - 09:51, Tuesday 09 August 2016 (28950)

9:30AM FRI AUG 5th, 2016 - Chamber entrance Particle counts with hand held CC counter:

Particle size Outside of chamber cover, in CR In chamber above table
0.3um 60 25
0.5um 20 10
1.0um 0 0

 

1:30PM FRI AUG 5th, 2016 - Chamber entrance Particle counts with hand held CC counter:

Particle size Outside of chamber cover, in CR In chamber above table
0.3um 60 0
0.5um 0 10
1.0um 0 0