WP 6073 I updated h0vaclx to change PT110 to a BPG 402 Inficon gauge. This required restarting the TwinCAT3 software and the EPICS IOC. I burtrestored the IOC to midnight yesterday. I also took the opportunity to enable and set the low trip point for PT110, PT170 and PT180 to 1E-5: (800E:02) Low Trip Point Enable TP1: TRUE (800E:14) Low Trip Point Limit TP1: 1E-5 (800F:02) Low Trip Point Enable TP2: TRUE (800F:14) Low Trip Point Limit TP2: 1E-5 The following channels were removed: H0:VAC-LX_Y0_110_PIRANI_INTLK H0:VAC-LX_Y0_110_PWR_REQ H0:VAC-LX_Y0_HS110_PWR_CTRL H0:VAC-LX_Y0_PT110A_PRESS_TORR H0:VAC-LX_Y0_PT110A_PRESS_TORR_ERROR H0:VAC-LX_Y0_PT110A_PRESS_VOLTS H0:VAC-LX_Y0_PT110A_PRESS_VOLTS_ERROR H0:VAC-LX_Y0_PT110A_PRESS_VOLTS_SMOO H0:VAC-LX_Y0_PT110B_PRESS_LOGTORR H0:VAC-LX_Y0_PT110B_PRESS_LOGTORR_ERROR H0:VAC-LX_Y0_PT110B_PRESS_TORR H0:VAC-LX_Y0_PT110B_PRESS_TORR_ERROR H0:VAC-LX_Y0_PT110B_PRESS_VOLTS H0:VAC-LX_Y0_PT110B_PRESS_VOLTS_ERROR The following channels were added: H0:VAC-LX_Y0_PT110_ERROR H0:VAC-LX_Y0_PT110_INFODATA_STATE H0:VAC-LX_Y0_PT110_MOD1_OVERRANGE H0:VAC-LX_Y0_PT110_MOD1_PRESS_TORR H0:VAC-LX_Y0_PT110_MOD1_PRESS_TORR_ERROR H0:VAC-LX_Y0_PT110_MOD1_SENSOR H0:VAC-LX_Y0_PT110_MOD1_TRIP H0:VAC-LX_Y0_PT110_MOD1_UNDERRANGE H0:VAC-LX_Y0_PT110_MOD1_VALID H0:VAC-LX_Y0_PT110_MOD2_OVERRANGE H0:VAC-LX_Y0_PT110_MOD2_PRESS_TORR H0:VAC-LX_Y0_PT110_MOD2_PRESS_TORR_ERROR H0:VAC-LX_Y0_PT110_MOD2_SENSOR H0:VAC-LX_Y0_PT110_MOD2_TRIP H0:VAC-LX_Y0_PT110_MOD2_UNDERRANGE H0:VAC-LX_Y0_PT110_MOD2_VALID H0:VAC-LX_Y0_PT110_WCSTATE_TOGGLE H0:VAC-LX_Y0_PT110_WCSTATE_WCSTATE