Krishna
Here are the recorded c-BRS associated channels and their calibrations (I think it is accurate to <20%):
1. H1:NGN-CS_CBRS_RY_PZT1_CTRL_IN1_DQ : PhotoDiode 1 output : 1.6e-10 rad/count
2. H1:NGN-CS_CBRS_RY_PZT1_MASTER_OUT_DQ : PZT1 output : 2.8e-10 rad/count
3. H1:NGN-CS_CBRS_RY_PZT2_CTRL_IN1_DQ : PhotoDiode 2 output : 1.23e-10 rad/count
4. H1:NGN-CS_CBRS_RY_PZT2_MASTER_OUT_DQ : PZT2 output : 2.8e-10 rad/count
I've also attached more spectra recorded this morning after the clean room fans were turned off. Many of the noise peaks went away, especially above 100 Hz, but some lines remain which are common to the seismometers and c-BRS. The DIFF signal above 100 Hz, drops to ~10 picorad/rt(Hz), indicating that it's noise floor is indeed well below the SUM channel. I expect that the current SNR at ~10 Hz is between 10-30. I've attached Matlab code showing the analysis.
Spectra from night time are attached.