Displaying report 1-1 of 1.
Reports until 11:39, Friday 19 August 2016
LHO VE
chandra.romel@LIGO.ORG - posted 11:39, Friday 19 August 2016 (29196)
HAM6 IP valved in; turbo valved out
Kyle, Chandra

Valved in IP and valved out TP. Pressure fell from 1.1e-6 Torr to 9e-7 Torr when IP was valved in and now is slowly rising with TP valved out (currently at 1.2e-6 Torr). After IP reaches thermal equilibrium we expect pressure to fall again. After that is observed we will shut down turbo + scroll (hopefully this afternoon).

Note: verified the MPC controller was ON and programmed for 500 L/s pump prior to valving back in. Read -5700 V. Now reads -5000 V with load.
Displaying report 1-1 of 1.