The power meter in the high power oscillator's external shutter was replaced. The existing unit ceased to function some time around the planned power outage a couple of months ago for reasons best known to itself. old S/N = 589365 new S/N = 627203 Jason/Peter
The hoses were swapped because we found that they were hooked up backwards. I.E. the MOPA cooling hose was plugged into the Power Meter water circuit and vice versa (most likely during the recent water manifold swap). This means that in the trend data the flow for the front end (H1:PSL-AMP_FLOW) was actually reading the flow through the power meter circuit; the same applies to the power meter circuit flow (H1:PSL-OSC_PWRMETERFLOW), it was actually reading the flow for the front end. This was fixed yesterday and the flow data is now reading from the correct water circuits.