Displaying report 1-1 of 1.
Reports until 16:59, Wednesday 24 August 2016
H1 ISC
terra.hardwick@LIGO.ORG - posted 16:59, Wednesday 24 August 2016 (29284)
Lockloss last night from known PI

We were locked last night at 50 W for ~2.5 hours and lost lock from an ITMX 15522 Hz PI. This is a known PI seen months ago; I had purposefully left damping settings off to see if it rang up. See first attachment showing lockloss

This afternoon we were locked at 50 W for ~2 hours and I let the mode ring up so I could demonstrate successful damping. See second attachment showing damping

All PIs that were previously observed have now been seen and damped post OMC vent. We have 5 PIs at 50 W  (at least up to ~3 hour locks): ITMX 15520 Hz; ETMX 15541 Hz; ETMY 15542 Hz, 15009 Hz, 18041 Hz (aliased from 47495 Hz).  All are successfully damped via the guardian and have had their damping phase and gain optimized. 

Images attached to this report