Jeff K, Darkhan T,
Last Tuesday we updated the infrastructure for injecting calibration lines (see LHO alogs 29245, 29249). Below is the table of currently active calibration lines:
Channel Names _FREQ (Hz) _SINGAIN (ct)
H1:CAL-PCALY_PCALOSC1_OSC 36.7 125 O1-scheme kappa_TST / kappa_PU
H1:CAL-PCALY_PCALOSC2_OSC 331.9 2900 O1-scheme kappa_C / f_C
H1:CAL-PCALY_PCALOSC3_OSC 1083.7 15000 high-frequency calibration check ("bonus" line)
H1:CAL-PCALX_PCALOSC1_OSC 3501.3 39322 high-frequency sensing function characterization ("mobile" line)
_FREQ (Hz) _CLKGAIN (ct)
H1:CAL-CS_TDEP_DARM_LINE1_DEMOD 37.3 0.1 O1-scheme kappa_PU
H1:SUS-ETMY_L3_CAL_LINE 35.9 0.11 O1-scheme kappa_TST / kappa_PU
H1:SUS-ETMY_L1_CAL_LINE 33.7 11 O2-scheme synched oscillator for kappa_U
H1:SUS-ETMY_L2_CAL_LINE 34.7 1.1 O2-scheme synched oscillator for kappa_P
H1:SUS-ETMY_L3_CAL2_LINE 35.3 0.11 O2-scheme synched oscillator for kappa_T
We plan to adjust the three O2-scheme line frequencies and amplitudes, and cancel them out with PCALY (they will not appear in the reconstructed DARM spectrum), following synchronized oscillators will be utilized for this purpose H1:CAL-PCALY_PCALOSC{4-6}_OSC.