Running a Worden experiment: -CP4 at 88% full -Set LLCV to nominal in manual mode (41% open) -Increase LLCV by 5% (43% open) and leave it at this value -Monitor increase in fill -Allow fill to exceed 100% and monitor exhaust flow meter (expect a sharp increase in flow when liquid starts to hit warm exhaust) -After this sharp increase, reduce LLCV by 10% to below nominal (<41%) -Hope that no LN2 reaches the flow meter this time (when flow approaches 100% I will monitor while sitting next to exhaust to be ready to open bypass exhaust valve and remove flow meter with LN2 gloves on) **THIS WILL GENERATE CP4 ALARM ON PUMP LEVEL**