Reports until 00:18, Saturday 03 September 2016
LHO General
thomas.shaffer@LIGO.ORG - posted 00:18, Saturday 03 September 2016 (29480)
Ops Eve Shift Summary

TITLE: 09/03 Eve Shift: 23:00-07:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Commissioning
INCOMING OPERATOR: None
SHIFT SUMMARY: Commissioning and an EQ. Toward the last bit of the night we made it to DC_READOUT_TRANSITION wher the OMC_LOCK Guardian went to the state, WAIT_FOR_AS_LIGHT. Seems that the light on the OMC QPDs was below its threshold of 100. Searching through alogs, Terra and I only found alog22289 that referenced this type of situation. We tried following Jenne's suggestion of making sure the Master gain was 0 (though it was actually 0.05, whoops) and then clearing the history of the ANG filter banks. But it didnt work, and when we tried to clear the history for the POS filter banks, because why not at this point, it just dropped the lock. We are done for the night, leaving it in down.