- The ISI configuration has been changed from 'Earthquake V2' to 'Windy'.

- Each time the ISC_LOCK guardian reaches CARM_15PM it stops and reports a connection error to H1:LSC-PD_DOF_MTRX_SETTING_1_23 (see attached screenshot). Manually running ezca on the guardian machine gets the value with no error. This can be gotten around by taking the node to manual and selecting CARM_10PM. 

- Making it through the DRMI ASC engagement is hit and miss even with careful alignment of the BS and PRM mirrors.

- I was not able to make it past ENGAGE_REFL_POP_WFS until Stefan came back. Stefan and Kiwamu found the DHARD ASC loop to be unstable.