P. King, J. Bartlett, J. Oberling
We swapped the PSL chillers for the backup in an effort to isolate the recent laser tripping issues. The laser is now back up and running.
Operating hours for the chillers, as reported by the chillers themselves and the chiller MEDM screen, are:
Chiller | MEDM | |
Crystal | 148.75 | 33317.00 |
Diode | 86.25 | 33560.00 |
There is a difference between the chillers and MEDM because the MEDM screen gets its info from the PSL Beckhoff PC, which uses its own internal counter to track the chiller hours instead of the hours recorded by the chillers themselves; we currently do not know a safe way to reset this Beckhoff counter.