Displaying report 1-1 of 1.
Reports until 00:35, Saturday 10 September 2016
H1 TCS (TCS)
matthew.evans@LIGO.ORG - posted 00:35, Saturday 10 September 2016 (29588)
ITMY ring heater increased to 2.5 W

Matt, Kiwamu

At around 0:30 local, I increased the ring heater power on ITMY from 1.5 W (0.75 W each segment) to 2.5 W (1.25 W each segment) (see 29557 for previous step). This has to be cancelled by applying approximately an additional 400 mW on CO2Y (e.g., pre-heat level of 800 or 900mW).  This step should allow us to better compensate the residual lens observed at 50W.

Displaying report 1-1 of 1.