Displaying report 1-1 of 1.
Reports until 09:55, Monday 12 September 2016
H1 TCS
nutsinee.kijbunchoo@LIGO.ORG - posted 09:55, Monday 12 September 2016 (29618)
TCS chillers water refilled + Note on power glitch

Happened on Saturday, forgot to log it.

450 ml for TCSX chiller

670 ml for TCSY chiller

Also cleaned both filters.

 

Note on power glitch: CO2 tripped as expected since h1oaf0 was down but h1hwsmsr computer didn't trip so both HWSX and HWSY codes were still running.

 

Side note: Matt mentioned that he couldn't find an instruction on how to restart the HWS code. That instruction can be found here.