Displaying report 1-1 of 1.
Reports until 13:13, Monday 12 September 2016
H1 ISC
sheila.dwyer@LIGO.ORG - posted 13:13, Monday 12 September 2016 (29625)
PR3 offset durring CARM offset reduction is back

Durring some of out unsucsesfull locking attempts last night we had the drift of PR3 durring the CARM offset reduction that we originally used the Ballmer offset to coutneract. (20055)

We had commented this out a few months ago because it was too large of a correction, but this morning I added it back and reduced it by a factor of 2 (the offset is now 0.25 in M1 P2P which causes about 0.02 urad of pitch).  