Reports until 11:44, Tuesday 13 September 2016
LHO VE
chandra.romel@LIGO.ORG - posted 11:44, Tuesday 13 September 2016 (29659)
CP4 overfill
Today CP4 reservoir overfilled to 100%+ after a Dewar fill because the lower limit of LLCV in PID settings was left at 37% from CP4 flow meter data collection. This was my mistake. All lower limits are normally set to 20%. I reset it to the default 20%. While I was at it, I lowered CP1 and CP2 lower limits to 10% since these two pumps normally run at a lower % open. 