Reports until 16:44, Tuesday 13 September 2016
H1 AOS
keita.kawabe@LIGO.ORG - posted 16:44, Tuesday 13 September 2016 (29676)
ISS second loop today (Daniel, Keita)

Some features were added in the model (first attachment):

Closed loop response of the AC coupling was calculated and a filter that mimics its behavior was put in FM6 of the third loop ("ACcpl") (second attachment bottom). Together with FM10 ("white") which mimics the analog whitening of the outer loop error point (which is basically zp=(78mHz^2,3.6^2)), this will make the outer loop transparent when viewed from the 3rd loop.

Images attached to this report