The CO2 heating power was calculated based on Aidan's CO2 power vs. PSL power plot (alog25932). With the new thermal lensing measurement (alog28799) I fine-tuned the equation.
CO2 power = slope * PSL power + offset
ITMX slope was -0.01, now -0.012
ITMY slope was -0.01, now -0.014
ITMX offset was 0.5, now 0.6
ITMY offset was 0.3, now 0.4
Corrected actuation plot. Divided lensing by a factor of 2 to make it a single-pass value.