A brief history of IP10, the controller for IP10 had A channel go somewhat bad on 9/3/2016, controller was set to output 5k volts but its output was only able to do about 2k, then we had a power glitch on 9/10/2016, that caused the current output to go down and the pressure to go up (see attached graph).
Since IP12 is currently valved out, where the pump is pumping on a small volume, we decided to swap controllers between pumps. So for now IP12 will show some alarm values until the controller is fixed. And to keep the small volume at IP12 good, we are pumping with one channel despite of what the signal says.