Reports until 16:36, Thursday 15 September 2016
H1 TCS
kiwamu.izumi@LIGO.ORG - posted 16:36, Thursday 15 September 2016 (29738)
HWSY complication

related to 29648,

The output from HWSY doesn't look reasonable. This is something we knew, but this time a surprising thing was that, if I interpret the data as it is, the self heating on ITMY introduced negative lensing rather than positive lensing. This does not make sense. See the second attachment below. The first attachment is the HWS signal for ITMX with its prediction from the TCS simulator for comparison.

Images attached to this report