Reports until 13:24, Monday 19 September 2016
H1 PSL
keita.kawabe@LIGO.ORG - posted 13:24, Monday 19 September 2016 (29798)
Second loop pointing monitor QPD dewhites were OFF

2nd loop QPD output alignment signals didn't make sense (attached top), and this turns out to be whitening mismatch.

This QPD goes through the QPD interface and standard whitening, and there is one unswitchable zp=0.4,40 analog whitening. This is supposed to be compensated by zp=(40,0.4) anti whitener (which is unfortunately named 0.4:40), which was off.

I turned them ON (attached left) and now it makes sense (bottom).

Images attached to this report