kyle.ryan@LIGO.ORG - posted 14:58, Monday 19 September 2016 (29803)
Manually over-filled Cryopump 3 (CP3)
LN2 at exhaust after 6 minutes 26 seconds of having the LLCV bypass valve opened 1/2 turn -> restored bypass valve to as found (closed) state.
Increased manual setting for LLCV from 16% to 17%