Jeff K, Dave B, Darkhan T,
Overview
Coherence calculation settings were updated today's after a minor bug fix in CAL-CS.
We discovered that some of the reference EPICS records that are used for calculation of kappas in the front-end were not set (or lost after a DAQ restart). The records were populated with values calculated from the DARM model for ER9.
Details
Similarly, SUS_LINE1 synched oscillator SINGAIN and COSGAIN values must be set to be $(IFO):SUS-ETMY_L3_CAL_LINE_CLKGAIN multiplied by DAQ downsampling TF magnitude at this frequency (35.9 Hz at LHO). Synched OSC phase must be set to the DAQ downsampling TF phase at the line frequency (not sure if we need to put positive or negative phase). Previously SUS_LINE1 synched OSC parameters were not set properly and the coherence was not calculated (see attachment 4, coherences calculated during last 4 days).
Both DARM_LINE1
and SUS_LINE1
oscillator parameters were updated, the changes were accepted in SDF_OVERVIEW.
H1:CAL-CS_TDEP_PCALY_LINE2_REF_A_USUM_(REAL|IMAG)
and H1:CAL-CS_TDEP_CAVITY_POLE_PCAL_FREQ
:
Old : H1:CAL-CS_TDEP_PCALY_LINE2_REF_A_USUM_REAL 0
New : H1:CAL-CS_TDEP_PCALY_LINE2_REF_A_USUM_REAL -1.73852e-19
Old : H1:CAL-CS_TDEP_PCALY_LINE2_REF_A_USUM_IMAG 0
New : H1:CAL-CS_TDEP_PCALY_LINE2_REF_A_USUM_IMAG -1.03541e-19
cds_user_apps
repo, r14297.