keita.kawabe@LIGO.ORG - posted 09:54, Wednesday 21 September 2016 - last comment - 00:12, Thursday 22 September 2016(29870)
ISS second loop OLTF at 50W

Attached is the OLTF in two configurations with 7dB VGA gain.

50W (actually 48W), AC coupling OFF, and boost ON (solid line)

50W (actually 48W), AC coupling ON, boost OFF (dashed).

UGF is about 9kHz with phase margin of 40 deg or so (boost ON) or 50 (boost OFF).

Boost buys us about 9dB more gain at 1kHz.

I asked Kiwamu to do an analog measurement up to higher frequency on the floor.

Images attached to this report
Comments related to this report
kiwamu.izumi@LIGO.ORG - 12:17, Wednesday 21 September 2016 (29877)

Here is a transfer fucntion with SR785 to check the high frequency portion.

Also, the attached zip contains the data, plot script, and figure.

Images attached to this comment
Non-image files attached to this comment
daniel.sigg@LIGO.ORG - 14:50, Wednesday 21 September 2016 (29884)

Taking this data I looked at modifications to the ISS board to increase its bandwidth. The attached plot shows:

  • brown — original measurement
  • green — adding a zero at 20kHz and adding a gain of 2
  • blue — same as green but also moving the knee of the boost from 2kHz to 4kHz
  • red —same as green but adding a second boost with knee at 2kHz.

Adding the 20kHz zero is a no brainer, since it only requires to limit the dominant pole in the circuit. It improves the phase margin over the measured curve and allows the ugf up to move to 20kHz. This will increase the ISS gain by 2 at 1kHz. The increased ugf and phase margin will in turn allow us to move the knee of the boost higher with the option to add a second boost. The later two solutions will add another gain of 2 at 1kHz.

Images attached to this comment
sheila.dwyer@LIGO.ORG - 00:12, Thursday 22 September 2016 (29899)

We engage the ISS boost twice today in low noise, and did not see any impact on the DARM noise. We expected an improvement at 1 kHz at least.  Attached are before and after spectra of the in and out of loop diodes (at least that's what I think the new cahnnel names mean).  If it is correct that INNER means in loop and outer means out of loop, we are sensor noise limited at 100 Hz with the boost on. 

Images attached to this comment