Kiwamu, Darkhan,
Summary
EPICS records used for calculation of the DARM time-dependent parameters have been updated with the values calculated using the most recent DARM model configuration parameters for ER10.
All κ's except for κPU calculated in the CAL-CS model are within reasonable ranges (see attachment 1):
κTST ≈ 0.75
κPU ≈ -0.25
κc ≈ 1.2
fc ≈ 345 Hz
We will investigate why κPU calculations are incorrect.
Details
The H1 DARM model parameters used for generation of the EPICS values are
Runs/PreER10/Common/params/IFOindepParams.conf
Runs/PreER10/H1/params/H1params.conf
Runs/PreER10/H1/params/H1params_2016-09-19.conf
Output files: raw EPICS values, a verbose log and a Matlab file with EP# variables (T1500377-08, Table 2) were committed to the calibration SVN (rev. 3335):
Runs/PreER10/H1/Scripts/CAL_EPICS
./callineParams_20160919.m
./20160921_H1_CAL_EPICS_VALUES.txt
./D20160921_H1_CAL_EPICS_VALUES.m
./20160921_H1_CAL_EPICS_verbose.log
./writeH1_CAL_EPICS.m
Runs/O2/Common/Scripts/CAL_EPICS/writeO2_TDEP_EPICS.m
The changes were accepted in SDF_OVERVIEW (safe.snap and OBSERVE.snap). For ER9 EPICS see LHO alogs 28180, 29104.