The diffracted power versus ISS offset slider value was measured. Attached is the plot for the percent diffracted power versus AOM drive signal. Unfortunately the plot isn't really parabolic. Looking at the plot it looks like the best region for the AOM drive is somewhere between 0.65 and 0.75 V. Which represents somewhere between 8.5 and 17%. A better characterisation of the AOM drive involves adjusting a pot on the ISS board to better linearise the output of the multiplier chip. That's a somewhat iterative process between measuring the diffracted output and adjusting the pot. The full power of the laser was measured to be 167.5 W.